Model
FD-R80
Supported pipe diameter
2 1/2″ (65 A), ø64 to ø83 2.52″ to 3.27″
3″ (80 A), ø83 to ø100 3.27″ to 3.94″
Supported pipe materials
Metal / resin*1
Supported fluids
Various liquids (i.e. water, oils, chemicals)*1
Supported fluid temperature
– 20 to + 120°C – 4.0 to + 248 °F (no freezing on the pipe surface)*2
Maximum rated flow
Rated flow velocity range: 5.0 m/s
Flow rate range (Typical):
2 1/2″ (65 A): 1000 L/min 260 gal/min 60 m3/ h)
3″ (80 A): 1500 L/min 390 gal/min 90 m3 / h
Zero cut flow rate
Zero cut: 0.3 m/s (default)*3
Flow rate (Typical): 90 L / min 24 gal / min 5.4 m3/h
Display method
Dual row, 5-digit display with white, 14-segment LED; Large status indicator; Output indicators; Stability indicator; Unit indicator
Display update cycle
Approx. 3 Hz
Display resolution
1 (L / min)
Response time
0.5 s / 1.0 s / 2.5 s / 5.0 s / 10.0 s / 30.0 s / 60.0 s / 120.0 s / 200.0 s (variable)
Measurement accuracy
Between 20 and 100% of F.S. : ±2.0% of RD*4*5
Between 6 and 20% of F.S. : ±0.4% of F.S. *4*5
Protection circuit
Power supply reverse connection protection, Power supply surge protection, Short-circuit protection for each output, Surge protection for each output
Zero point error
±0.5% of F.S.*4*6
Hysteresis
Variable
Flow units
L / min, m3 / h, gal / min
Integrated flow unit display
1 / 10 / 100 / 1000 / 10000 (L)
Pipe temperature measurement accuracy
±3°C ±5.4°F (liquid temperature of – 20 to + 50°C, – 4 to + 122°F)
±5°C ±9°F (liquid temperature of 50 to + 120°C, 122 to 248°F)
(ambient operating temperature of 25°C 77°F)*4
Wiring specifications
Power supply
DC power supply: M12 4-pin connector / AC power supply: M4 screw terminal block (selectable)
I/O
When using a DC power supply: M12 4-pin connector / when using an AC power supply: M3 screw terminal block
Input/Output (Selectable)
Output (ch.1/ch.2)
Control output / Integrated pulse output / Error output / Temperature alarm, NPN / PNP setting switchable,open collector output 30 VDC or less, max. 100 mA / ch., residual voltage: 2.5 V or less*7
Analog output (ch.1/ch.2)
Flow rate analog output / Temperature analog output and 4 – 20 mA / 0 – 20 mA (selectable), load resistance: 500 Ω or less*7
External input (ch.2)
Integrated flow reset input / Flow rate zero input / Origin adjustment input (selectable), short-circuit current: 1.5 mA or less, input time: 20 ms or more*7
Rating
Power voltage
20 to 30 VDC including 10% ripple (P-P), Class 2 / 100 to 240 VAC – 15% or + 10% (50 / 60 Hz)
Current consumption
When using a DC power supply: 200 mA or less (load current excluded), 400 mA or less (load current included) When using an AC power supply: 15 VA or less
Environmental resistance
Enclosure rating
IP65 / 67 (IEC60259), IP69K (ISO20653), Enclosure Type 4X (NEMA250)
Ambient temperature
– 20 to + 60°C – 4.0 to 140 °F (no freezing)*2
Relative humidity
5 to 90%RH (no condensation)
Vibration resistance
10 to 55 Hz, compound amplitude 1.5 mm 0.06″, XYZ axes 2 hours for each axis
Shock resistance
100 m/s2 , 16 ms pulse, XYZ axes, 1000 times for each axis
Dimensions
Main unit size: 218.5 mm × 66.9 mm × 70.7 mm 8.60″ × 2.63″ × 2.78″
Material
Main unit
Body: aluminum die-casting + coating / PPS, display: reinforced glass, connectors: SUS304-equivalent
Unit rear
Rubber
Upper/lower bracket
SUS304
Weight
Main unit: Approx. 1.0 kg
Upper/lower bracket: Approx.2.0 kg (including sub unit)
*1 Liquid must allow for the passage of an ultrasonic pulse, as well as not contain large air pockets or excessive bubbles. Detection may be unstable due to the type and status of the pipes.
*2 Perform derating depending on the ambient temperature and liquid temperature when using an AC power supply.
*3 The zero cut flow rate can be changed in the settings.
*4 This value is guaranteed by KEYENCE inspection facilities. Errors will be introduced by the type and status of the pipes, the type and temperature of the fluid, and the zero cut flow rate.
*5 This is the value when considering linearity + span error + repeatability in a stable environment of 25°C 77°F.
*6 It is possible to enhance the precision of zero point error by performing an origin adjustment.
*7 IO-Link: Compatible with Specification v1.1 / COM2 (38.4 kbps) The setting file can be downloaded from the KEYENCE website. If using the unit in an environment where downloading the file is not accessible via Internet, contact your nearest KEYENCE office. IO-Link is either registered trademarks or trademarks of PROFIBUS Nutzerorganisation e.V. (PNO)