Product characteristics
Number of inputs and outputs
Number of analogue outputs: 2
Measuring range
0…200 l/min
0…52.84 gpm
Process connection
threaded connection G 1 1/4 flat seal
Application
Special feature
Gold-plated contacts
Application
for industrial applications
Installation
connection to pipe by means of an adapter
Media
water; glycol solutions; coolants; oil
Note on media
low-viscosity oils with viscosity: 7…40 mm²/s (40 °C)
high-viscosity oils with viscosity: 30…68 mm²/s (40 °C)
Medium temperature [°C] -10…80
Pressure rating [bar] 16
Pressure rating [Mpa] 1.6
Electrical data
Operating voltage [V] 19…30 DC; (according to EN 50178 SELV/PELV)
Current consumption [mA] 100
Min. insulation resistance [MΩ] 100; (500 V DC)
Protection class
III
Reverse polarity protection
yes
Power-on delay time [s] 10
Inputs / outputs
Number of inputs and outputs
Number of analogue outputs: 2
Outputs
Total number of outputs
2
Output signal
analogue signal
Number of analogue outputs
2
Analogue current output [mA] 4…20; (scalable)
Max. load [Ω] 500
Short-circuit protection
yes
Type of short-circuit protection
pulsed
Overload protection
yes
Measuring/setting range
Measuring range
0…200 l/min
0…52.84 gpm
Display range
0…240 l/min
0…63.42 gpm
Resolution
0.1 l/min
0.02 gpm
Analogue start point ASP
0…160 l/min
0…42.28 gpm
Analogue end point AEP
40…200 l/min
10.28…52.84 gpm
Max. flow rate [l/min] 220
In steps of
0.1 l/min
0.02 gpm
Temperature monitoring
Measuring range [°C] -10…80
Resolution [°C] 0.2
Analogue start point [°C] -10…62
Analogue end point [°C] 8…80
In steps of [°C] 0.2
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring range)
20 l/min)
Response times
Flow monitoring
Response time [s] 0.25; (dAP = 0)
Damping for the switching output dAP [s] 0…1
Temperature monitoring
Dynamic response T05 / T09 [s] T09 = 30 (Q > 20 l/min); (water)
Operating conditions
Ambient temperature [°C] -10…60
Storage temperature [°C] -25…80
Protection
IP 67
Mechanical data
Weight [g] 1883
Materials
housing: AlMgSi0,5 anodised; Sealing: FKM; connector housing: brass Optalloy-plated; PA 6.6; cover film: PA
Materials (wetted parts)
stainless steel (1.4404 / 316L); FKM; PPS; Centellen 200
Process connection
threaded connection G 1 1/4 flat seal
Displays / operating elements
Display
Display unit
6 x LED, green (l/min, m³/h, gpm, gph, °C, °F)
measured values
alphanumeric display, 4-digit
programming
alphanumeric display, 4-digit
Display unit
l/min; m³/h; gpm; gph; °C; °F
Accessories
Accessories (supplied)
sealings: 2, Centellen
Accessories (optional)
adapter for pipe: 1 x R 1, stainless steel, E40205
adapter for pipe: 1 x 1″ NPT, stainless steel, E40206
Remarks
Remarks
sealing: only with supplied Centellen seals
MW = measured value
MEW = Final value of the measuring range
Pack quantity
1 pcs.
Electrical connection
Connection